Common faults

皓辰团队专业维修,品质保障!!!
专业维修奥林巴斯 BX51M 金相显微镜,其常见的故障及原因分析如下
故障点:照明亮度调亮后,视场亮度变化不大
原因分析:光源中心偏离、瞳距调整不正确、屈光度调整不正确、左右目镜放大倍数不同。
故障点:转动粗准焦螺旋时,镜筒不能升降
原因分析:齿轮与齿条未吻合,可能是齿杆套未固定好或齿条磨损。
故障点:焦距对准后,镜筒自行下滑
原因分析:垫圈磨损变形,导致齿轮轴与齿杆套间的摩擦力减少。
故障点:光柱不能定位
原因分析:集光器的滚珠遗失或弹簧失效。
故障点:光圈无法改变
原因分析:遮光片上的小钢柱脱落。
故障点:镜臂不能随意倾斜
原因分析:倾斜关节松动或摩擦垫圈磨损。
通过了解设备的常见故障,您可以更好地维护和使用它,确保其长期稳定运行。若无法解决,欢迎联系晧辰!我们的维修团队拥有丰富的实战经验和专业的维修流程,可为您提供快速、可靠、优质的维修服务!

Maintenance case

Hao Chen team professional maintenance, quality assurance!!

Professional maintenance of Olympus BX51M metallographic microscope, its common failures and reasons are analyzed as follows

Fault point: After the lighting brightness is brightened, the field of view brightness changes little

Cause analysis: light source center deviation, pupil distance adjustment is not correct, diopter adjustment is not correct, left and right eyepiece magnification is different.

Fault point: When rotating the coarse primary focus spiral, the mirror can not rise and fall

Cause analysis: the gear and rack are not matched, which may be that the gear rod sleeve is not fixed or the rack is worn.

Fault point: After the focal length is aligned, the lens barrel slides down by itself

Cause analysis: Gasket wear deformation, resulting in reduced friction between the gear shaft and the gear rod suite.

Fault point: Light column can not be located

Cause analysis: light collector ball loss or spring failure.

Failure point: The aperture cannot be changed

Cause analysis: The small steel column on the light shield falls off.

Fault point: The mirror arm can not tilt at will

Cause analysis: inclined joint loosening or friction washer wear.

By understanding the common faults of the equipment, you can better maintain and use it to ensure its long-term stable operation. If you can not solve, welcome to contact Hao Chen! Our maintenance team has a wealth of practical experience and professional maintenance process, can provide you with fast, reliable, quality maintenance service!

Industry classification

1070558

Similar hot instrument

Friendship recommendation

Cooperative client:

more>>

Domestic instrument:

more>>

Import instrument:

more>>

Metrology calibration:

more>>

Copyright © Copyright © 2007 All Rights Reserved. Copyright© 2003-2015 Shenzhen Haochen Electronic Technology Co.,ltd . All Rights Reserved.   [粤ICP备14101712号-1] [Legal Adviser]

Address:8th Floor, NO.221, Niucheng Road, Xili Town,Nanshan District, Shenzhen, Guangdong Province, China,518055

QQ online service:  Service hotline:0755-86016691  Tel:13715327187  Fax:+86 0755-86641139-816

E-mail:info@wl95.com